![]() MARD was 15.5% for glucose values 14 mmol/L. ![]() Overall MARD was 11.5 (95% CI, 10.7-12.3)% with 99% of readings in Clarke zones A and B. We used mean absolute relative difference (MARD) and Clarke error grid analysis to assess accuracy of matched CGM-blood glucose values overall, across glucose stata (14 mmol/L), and over time (≤48, 48-96, >96 h). We used linear regression analysis adjusted for illness severity to assess the association of CGM use with blood gas sampling frequency and glucose variability. Both groups received insulin to target glucose range of 10-14 mmol/L. We used the FreeStyle Libre CGM in 15 ICU patients and compared their blood glucose metrics with a pre-CGM control population of 105 ICU patients with type 2 diabetes. We aimed to compare CGM to blood gas glucose values and assess whether CGM use reduces blood gas sampling frequency and glucose variability in ICU patients with type 2 diabetes managed with liberal glucose control. Whether subcutaneous continuous glucose monitoring (CGM) can safely replace intermittent arterial blood gas glucose analyses in intensive care unit (ICU) patients remains uncertain.